In-Situ Inc.
221 E Lincoln Ave.
Ft. Collins, CO 80524
970-498-1500

Multi-Parameter TROLL® 9500

Item

Specification

Accuracy and precision

+/- 0.1 mg/L from 0-8 mg/L; +/-0.2mg/L from 8-20 mg/L

Resolution

0.01 mg/L

Operating range

0-20 mg/L or 0-450% saturation

Temperature range (operating)

0 to 50° C (32 to 122° F)

Pressure rating

300 PSI (210 m; 692 ft)

Salinity range

Up to 42 ppt

pH range

2-10 pH

Typical response limit

>25 mg/L

Response time

T90 = 30 seconds; T95 = 37 seconds

Technology

Lumiphore impregnated foil matrix, dynamic luminescence quenching technique using phase domain detection.

Calibration

1 or 2 point

Typical calibration duration

Recommended calibration check every 3 months (clean water)

Known interferences

Interferences (cross-sensitivity) are found for alcohols > 5%; hydrogen peroxide (H2O2) > 3%; sodium hypochlorite (NaClO; commercial bleach) > 3%; gaseous sulfur dioxide (SO2); and gaseous chlorine (Cl2).

The In-Situ® optical DO sensor should not be used in organic solvents, such as acetone, chloroform or methylene chloride, which may swell the foil matrix and destroy it.

No cross sensitivities exist for carbon dioxide (CO2), ammonia (NH3), pH, any ionic species like sulfide (S2- ), sulfate (SO42-), chloride (Cl-), or hydrogen sulfide (HS-).

Material

Delrin®

Typical sensor foil life

1 year or 1 million measurements

Cap usage life

1 year from the first TROLL 9500 instrument reading. For full operational use of the RDO cap, install by date printed on package.

Dimensions, RDO® sensor

25.4 mm (1 in) OD, 14 cm (5.5 in) long

Dimensions, RDO® sensor adapter assembly

88.4 mm (3.5 in) OD, 20.3 cm (8.0 in) long

Storage conditions, sensor

-5 to 60° C (23 to 140° F)

Storage conditions, cap

Prior to installation: 1 to 60° C (33 to 140° F); store in factory supplied container

Warranty, sensor

3 years from date of manufacture

Warranty, sensor cap

2 years from date of manufacture